Skip to main content
The Medical Device Safety Solution implements this decision with dedicated fields on each risk record, a structured evaluation workflow, and a final risk determination that auditors can trace from hazard to disposition.

When Benefit-Risk Analysis Is Required

Benefit-risk analysis applies to a specific subset of risk records: diagram The three post-mitigation outcomes:
Post-Risk LevelBenefit-Risk Required?Next Step
AcceptableNoRecord final risk as Acceptable
InvestigationYesPerform benefit-risk analysis
UnacceptableNo (but action required)Apply additional risk controls or redesign

The ISO 14971 Decision Sequence

ISO 14971 defines a specific sequence for residual risk that is in the Investigation zone:
  1. Determine if additional controls are practicable — Can the risk be reduced further through design, protective measures, or information?
  2. If additional controls exist — Apply them and re-evaluate. Repeat until no further reduction is practicable.
  3. If no additional controls are practicable — Perform benefit-risk analysis.
  4. Document the benefit — What clinical or patient benefit does the device provide that justifies accepting the residual risk?
  5. Evaluate the balance — Does the benefit outweigh the risk?
The solution captures this sequence through four fields on each risk record:
FieldIDTypePurpose
Additional Controls PossibleadditionalControlsPossibleYes/No enumStep 1-2: Are further controls practicable?
BenefitbenefitRich textStep 4: Clinical benefit description
Benefit-Risk ResultriskBenefitResultEnumStep 5: Benefit > Risk (acceptable) or Benefit < Risk (not acceptable)
Final RiskfinalRiskEnumConclusive disposition

Benefit-Risk Result Values

The riskBenefitResult enumeration has two values:
ValueDisplay NameMeaning
benefitAcceptableBenefit > Risk — acceptableThe clinical benefit outweighs the residual risk
benefitNotAcceptableBenefit < Risk — NOT acceptableThe residual risk exceeds the clinical benefit
If the benefit-risk result is NOT acceptable, the device cannot be released with this risk record in its current state. Either additional risk controls must be found, the design must be changed, or the intended use must be modified to reduce the risk.

Final Risk Determination

The finalRisk enumeration captures the conclusive risk disposition. Unlike preRisk and postRisk (which have the Investigation option), final risk forces a definitive decision:
ValueDisplay NameMeaning
acceptableAcceptableRisk reduced to a tolerable level
unacceptableUnacceptableRisk remains too high despite all measures
benefitAcceptableBenefit-Risk AcceptableResidual risk accepted because benefits outweigh it
The absence of an “Investigation” option in final risk is deliberate. By the time a risk record reaches final determination, the team must have made a definitive judgment. There is no “still investigating” state for released products.

Practical Example

Consider a risk record for the Smart Infusion Pump:
StepFieldValue
HazardAir bubble in fluid path
HarmAir embolism (Severity: Critical)
Pre-mitigation riskpreRiskUnacceptable
Risk control appliedUltrasonic bubble detector with auto-shutoff (Protective Measure)
Post-mitigation riskpostRiskInvestigation
Additional controls?additionalControlsPossibleNo
Benefitbenefit”The infusion pump enables precise, continuous medication delivery to critically ill patients, replacing manual IV administration that has higher error rates and cannot maintain constant flow rates. Without this device class, patients requiring vasoactive medications face significantly higher mortality risk from dosing errors.”
Benefit-Risk ResultriskBenefitResultBenefit > Risk — acceptable
Final RiskfinalRiskBenefit-Risk Acceptable

The ALARP Connection

The Investigation zone in the risk matrix corresponds to the ALARP (As Low As Reasonably Practicable) principle. Risks in this zone are neither clearly acceptable nor clearly unacceptable. The benefit-risk analysis provides the documented justification for accepting residual risk in this zone. ALARP requires demonstrating that:
  • All reasonably practicable risk controls have been applied
  • The cost of further risk reduction is grossly disproportionate to the benefit
  • The clinical benefit of the device justifies the remaining risk
The additionalControlsPossible field captures the first criterion. The benefit field and riskBenefitResult capture the second and third.

Audit Trail

For ISO 14971 compliance, the benefit-risk analysis must be traceable. The solution provides this through:
  • Risk Record fields — The four benefit-risk fields create a structured audit trail on each record
  • Risk Control Plan links — Each risk control that was applied before the benefit-risk analysis is linked and classified
  • Document workflow — The HARA document goes through formal review and approval, capturing electronic signatures
  • Risksheet history — Changes to risk fields are tracked through Polarion’s work item history